R&S
®
ZVA/ZVB/ZVT SYSTem...
1145.1084.12 6.289 E-6
SYSTem:COMMunicate:RDEVice:GENerator<gen_no>:DEFine '<gen_name>',
'<driver>', '<interface>', '<address>'[, <fast_sweep>, <10_MHz_Ref>]
Configures an external generator and adds it to the list of available generators.
<gen_no>
Number of the configured generator. Generators must be numbered
in ascending order, starting with 1. If a number is re-used, the
previous generator configuration is overwritten.
Generators can be assigned several times so that the number of
configured generators is practically unlimited.
'<gen_name>'
Name of the external generator (string parameter). An empty string
means that no particular name is assigned to the generator.
'<driver>'
Generator type (string parameter). The generator type is identical
with the name of the generator driver file (*.gen) stored in the
resources\extdev subdirectory of the analyzer's program directory.
Alternative: The '<gen_name>' string may contain the generator
driver file name with its complete path, e.g. 'C:\Program
Files\Rohde&Schwarz\Network
Analyzer\resources\extdev\sme02.gen'.
'<interface>'
Interface type (string parameter): 'GPIB0', 'VXI-11',...
'<address>'
Interface address (string parameter), depending on the interface
type.
<fast_sweep>
Optional Boolean parameter, enables or disables the fast sweep
mode.
<10_MHz_Ref>
Optional Boolean parameter, sets the analyzer to internal (OFF) or
external (ON) reference frequency.
*RST values *RST has no effect on the configured generators.
SCPI,
Command
Types
Device-specific, command or query
Example:
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02', 'gpib0', '21'
Configure an R&S SNE02 generator as external generator no. 1,
assigning the name Ext. Gen. 1 and a GPIB address 21.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The response 'Ext. Gen.
1','SME02','GPIB0','21',OFF, OFF indicates that the fast
sweep mode is disabled and that the analyzer is set to internal
reference frequency.
SYST:COMM:RDEV:GEN1:DEL
Clear the generator configuration table.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The analyzer returns an error
message because the generator no. 1 is no longer configured.
SYSTem:COMMunicate:RDEVice:GENerator<gen_no>:DELete
Clears the configuration table for external generators.
<gen_no>
Number of the configured generator. This number is ignored, the
command clears all entries in the configuration table.