EasyManua.ls Logo

Rohde & Schwarz R&S ZVA 8 - Page 757

Rohde & Schwarz R&S ZVA 8
813 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S
®
ZVA/ZVB/ZVT SYSTem...
1145.1084.12 6.295 E-6
SCPI, Command
Types
Confirmed, no query
Example:
INIT:CONT OFF
Activate single sweep mode.
SYST:SETT:UPD ONCE
Update the settings made during the current single sweep
period.
SYSTem:SOUNd:ALARm[:STATe] <Boolean>
Switches alarm sounds on or off.
<Boolean>
OFF - Alarm sounds switched off
ON - Alarm sounds switched on
*RST value ON
SCPI, Command Types
Device-specific, command or query.
Example:
SYST:SOUN:ALAR OFF; STAT OFF
Switch alarm and status sounds off.
SYSTem:SOUNd:STATus[:STATe] <Boolean>
Switches status sounds on or off.
<Boolean>
OFF - Status sounds switched off
ON - Status sounds switched on
*RST value ON
SCPI, Command Types
Device-specific, command or query.
Example:
See SYSTem:SOUNd:ALARm[:STATe]
SYSTem:VERSion?
Returns the SCPI version number to which the analyzer complies. The analyzer complies to the final
SCPI version 1999.0
*RST value
SCPI, Command Types
Confirmed, query only
Example:
SYST:VERS?
Query the SCPI version. The response is 1999.0.