Data Logging
R&S
®
Scope Rider RTH
163User Manual 1326.1578.02 ─ 12
Figure 8-4: Cursor settings
Enable
Enables or disables the cursor measurement.
Remote command:
LOGGer:CURSor[:STATe] on page 418
Cursor Value
Sets the measured crossing point between the cursor lines and the waveform, see Fig-
ure 8-5. These settings only apply if data logging is running for more than 2 days and 7
hours. After this time, the logger compresses 4 logging values into a "Minimum", "Aver-
age" and "Maximum" value.
Figure 8-5: Logger cursor with summarized data
1 = Maximum value of cursor line 1
2 = Maximum value of cursor line 2
Remote command:
LOGGer:CURSor:TYPE on page 418