Remote Control Commands
R&S
®
Scope Rider RTH
370User Manual 1326.1578.02 ─ 12
HARMonic:LIMits:CURRent?
Return values:
<Current>
Usage: Query only
HARMonic:LIMits:TYPE <Type>
Defines the type of limit check to be performed.
Parameters:
<Type> NONE | EN50160 | USER
NONE
No limit check is performed.
EN50160
Limits are checked according to the predefined values in the
EN50160 standard.
USER
Limits are checked according to the values in a user-defined file
(see HARMonic:LIMits:LOAD on page 369). For details on
the required file format see Chapter 6.3.6.1, "Limit File Format",
on page 141.
*RST:
NONE
HARMonic:SCALe <Unit>
Switches the scale for the measured harmonic levels between logarithmic (db) and lin-
ear (percent). The values are relative to the level measured for the fundamental fre-
quency, or to the level measured for the entire signal, depending on the selected THD
type (see HARMonic:THDType on page 371).
Parameters:
<Unit> PERCent | DB
*RST: DB
HARMonic:STATistic <Type>
By default, the numeric results indicate the currently measured values. Optionally, you
can switch to the minimum or maximum values.
Parameters:
<Type> CURRent | MIN | MAX
*RST: CURRent
Spectrum Analysis