Remote Control Commands
R&S
®
Scope Rider RTH
381User Manual 1326.1578.02 ─ 12
Parameters:
<Range> <numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 A|10 A|100 A|
1000 A.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 1 to 1000
*RST: 1
Default unit: A
Usage: Setting only
METer:CONFigure:FREQuency
Configures the frequency measurement and sets voltage autoranging.
To set a fixed voltage range, use METer:SENSe:FREQuency:VOLTage:RANGe:
UPPer.
Usage:
Event
METer:CONFigure:RESistance
Sets the internal parameters and configures the range for resistance measurements.
Parameters:
<Range> <numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 k|10 k|
100 k|1 M|10 M|100 M.
You can enter any value between the minimum and maximum
value. The instrument sets the next suitable range.
Range: 1 kOhm to 100 MOhm
*RST: AUTO
Usage: Setting only
METer:CONFigure:VOLTage:AC
METer:CONFigure:VOLTage:DC
Sets the internal parameters and configures the range for voltage measurements.