Remote Control Commands
R&S
®
Scope Rider RTH
445User Manual 1326.1578.02 ─ 12
BUS:UART:TECHnology <ThresholdType>
Sets the threshold value of the UART line as defined for various signal technologies.
Parameters:
<ThresholdType> TTL | ECL | CMOS | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
USER
Set a user-defined value using BUS:UART:THReshold.
*RST: TTL
BUS:UART:THReshold
<ThresholdValue>
Sets an individual threshold value for digitization of signals if BUS:UART:TECHnology
is set to USER.
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
BUS:SETReflevels
Sets the appropriate threshold or threshold for the selected serial protocol.
Usage:
Event
BUS:UART:STDBitrate <Bitrate>
Sets the number of transmitted bits per second.
Parameters:
<Bitrate> BPS_300 | BPS_600 | BPS_1200 | BPS_2400 | BPS_4800 |
BPS_9600 | BPS_14400 | BPS_19200 | BPS_28800 |
BPS_38400 | BPS_56000 | BPS_57600 | BPS_115200 |
BPS_128000 | BPS_230400 | BPS_460800 | BPS_921600 |
CUSTom
Values in bits per second.
CUSTom: Set the bit rate using BUS:UART:BITRate.
*RST: BPS_14400