Remote Control Commands
R&S
®
Scope Rider RTH
475User Manual 1326.1578.02 ─ 12
BUS:SENT:CRCMethod..................................................................................................476
BUS:SENT:CRCVersion..................................................................................................476
BUS:SENT:DNIBbles......................................................................................................476
BUS:SENT:PPFLength................................................................................................... 477
BUS:SENT:PPULse........................................................................................................477
BUS:SENT:SFORmat..................................................................................................... 477
BUS:SENT:RDSL...........................................................................................................477
BUS:SENT:DATA:SOURce <Channel>
Sets the source of the data line. All channel waveforms can be used.
Parameters:
<Channel> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
C3 and C4: only R&S RTH1004
Digital channels are available if option R&S RTH-B1 is installed.
*RST: C1
Usage: Asynchronous command
BUS:SENT:POLarity <Polarity>
Sets the idle state.
Parameters:
<Polarity> IDLLow | IDLHigh
*RST: IDLHigh
BUS:SENT:TECHnology <ThresholdType>
Sets the threshold values as defined for various signal technologies.
Parameters:
<ThresholdType> CMOS | USER
*RST: USER
BUS:SENT:DATA:THReshold <ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for BUS:SENT:TECHnology on page 475 USER.
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
Protocol Analysis