Remote Control Commands
R&S
®
Scope Rider RTH
481User Manual 1326.1578.02 ─ 12
BUS:SENT:FRAMe<m>:NIBBle<n>:STATe?......................................................................482
BUS:SENT:FRAMe<m>:NIBBle<n>:VALue?..................................................................... 483
BUS:SENT:FRAMe<m>:PAPTicks?..................................................................................483
BUS:SENT:FRAMe<m>:SCOM?......................................................................................483
BUS:SENT:FRAMe<m>:SDATa?......................................................................................484
BUS:SENT:FRAMe<m>:STARt?...................................................................................... 484
BUS:SENT:FRAMe<m>:STATus?.................................................................................... 484
BUS:SENT:FRAMe<m>:STOP?.......................................................................................485
BUS:SENT:FRAMe<m>:SYMBol?....................................................................................485
BUS:SENT:FRAMe<m>:SYNCduration?...........................................................................485
BUS:SENT:FRAMe<m>:TYPE?.......................................................................................485
BUS:SENT:FCOunt?
Returns the number of decoded frames.
Return values:
<Count> Total number of decoded frames.
Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS:SENT:FRAMe<m>:CSValue?
Returns the CRC sequence value of the selected frame.
Suffix:
<m>
.
*
Return values:
<ChecksumValue> The values below – range, increment and reset – are decimal
values.
Range: 0 to 63
Increment: 1
*RST: 0
Usage: Query only
BUS:SENT:FRAMe<m>:DATA?
Returns the data of the specified frame.
Suffix:
<m>
.
*
Example:
BUS:SENT:FRAMe4:DATA?
<-- 4,3,15,11,9
Usage: Query only
Protocol Analysis