Waveform Setup
R&S
®
Scope Rider RTH
68User Manual 1326.1578.02 ─ 12
Remote command:
TRIGger:LEVel<m>:SLEW:UPPer on page 325
TRIGger:LEVel<m>:SLEW:LOWer on page 325
Slope
Sets the edge direction for the trigger. You can trigger on:
rising edge, that is a positive voltage change
falling edge, that is a negative voltage change
rising and falling edge
Remote command:
TRIGger:SLEW:SLOPe on page 325
Range
Defines the time limits of the slew rate. The comparisons are the same as for the width
trigger, see "Range" on page 56.
Remote command:
TRIGger:SLEW:RANGe on page 325
Time
Sets the slew rate for comparisons equal, unequal, shorter, and longer.
Remote command:
TRIGger:SLEW:RATE on page 325
±Tolerance
Sets a tolerance range Δt to the specified Time if the comparison range is equal or
unequal.
Remote command:
TRIGger:SLEW:DELTa on page 326
Min Time / Max Time
Set the lower and upper time limits if "Inside" or "Outside" is set for comparison.
Remote command:
TRIGger:SLEW:MINWidth on page 326
TRIGger:SLEW:MAXWidth on page 326
3.6.11 Data2Clock Trigger (R&S RTH-K19)
With the Data2Clock trigger - also known as setup/hold trigger - you can analyze the
relative timing between two signals: a data signal and the synchronous clock signal.
Many systems require, that the data signal must be steady for some time before and
after the clock edge, for example, the data transmission on parallel interfaces.
The reference point for the time measurement is defined by clock level and clock edge.