User Correction R&S SMB
1407.0806.32 4.40 E-2
State - User Correction
Activates/deactivates user correction.
The UCOR status message appears in the frequency and level
display.
Remote-control command:
SOUR:CORR:STAT ON
Value - User Correction
Indicates the current value for level correction.
Remote-control command:
SOUR:CORR:VAL?
User Correction Data - User
Correction
Calls the File Select menu for selecting and creating a list or the File
Manager.
Remote-control command:
SOUR:CORR:CSET:CAT?
Response:
'Ucor1', 'Ucor2'
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DEL 'Ucor2'
Edit User Correction Data -
User Correction
Calls the editor for editing the selected user correction list. A list
consists of any number of frequency/level value pairs. The currently
selected list is displayed.
Important:
Save list only after filling both columns (frequency and level),
otherwise the entries are lost.
Frequency /Hz
Enter the frequency to which the level correction
value applies.
Remote-control command:
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DATA:FREQ 100MHz,...
Power /dB
Enters the level correction value to which the
specified frequency applies.
Remote-control command:
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DATA:POW 1dB,0.8dB,...