Sweep Mode R&S SMB
1407.0806.32 4.46 E-2
Single
Sets a single sweep cycle. The sweep is triggered
by the Execute Single Sweep button.
If a different sweep mode was activated prior to the
Single mode, the current sweep is stopped. The
Single sweep always starts at the start frequency
when triggered. The Reset button stops the sweep
and resets it to the start frequency.
Remote-control commands:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
Step
Sets a step-by-step sweep cycle.
If this mode is activated, the cursor moves to the
value displayed for Current Freq. Any variation to
the Current Freq value triggers a sweep step. The
step width is set below at entry field Step Lin or
Step Log.
If a different sweep mode was activated prior to the
Step mode, the current sweep is stopped. The Step
sweep starts at the current RF frequency.
The Reset button resets the sweep to the start
frequency.
Remote-control commands:
SOUR:SWE:FREQ:MODE MAN
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:SPAC LIN
SOUR:SWE:FREQ:STEP:LIN 0.5E4
SOUR:FREQ:MAN 1E6
(the value entered with command SOUR:SWE:
FREQ:STEP:LIN|LOG sets the step width. The
value entered with command SOUR:FREQ:MAN has
no effect, the command only triggers the next sweep
step. However, the value has to be in the currently
set sweep range (start to stop). In remote control
only a step-by-step sweep from start to stop
frequency is possible)
Extern Single
Sets a single sweep cycle. The sweep is triggered
by an external trigger signal.
If a different sweep mode was activated prior to the
Extern Single mode, the current sweep is stopped.
The Extern Single sweep always starts at the start
frequency when triggered. The Reset button resets
the sweep to the start frequency.
The external trigger signal is input at the rear of the
instrument (BNC connector INST TRIG).
Remote-control command:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR EXT
SOUR:FREQ:MODE SWE
(External trigger)