R&S SMB LF Generator and LF Output - Mod Gen Block
1407.0806.32 4.79 E-2
Extern
Start/Stop
Sets an automaticly repeated sweep cycle that is
started, stopped and restartet by subsequent
external trigger events.
The first external trigger signal starts the sweep
(Start).
The next external trigger signal stops the sweep at
the current frequency (Stop).
The third external trigger signal starts the sweep at
the start frequency (Start).
If a different sweep mode was activated prior to the
Extern Start/Stop mode, the current sweep is
stopped and the Extern Start/Stop sweep starts at
the start frequency when triggered.
The Reset button resets the sweep to the start
frequency.
The external trigger signal is input at the rear of the
instrument (BNC connector INST TRIG).
Remote-control command:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SOUR EAUT
SOUR:LFO:FREQ:MODE SWE
(External trigger)
Reset Sweep - LF Sweep
Resets the sweep. The start frequency is set and the next sweep
starts from there.
Remote-control command:
SWE:RES:ALL
Execute Single Sweep - LF
Sweep
Triggers the sweep manually. A manual sweep can only be triggered if
Mode Single is selected.
Remote-control commands
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR SING
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:EXEC