R&S SMB Command Description
1407.0806.32 6.9 E-2
DIAGnostic:INFO:OTIMe?
The command queries the number of operation hours
Example:
DIAG:INFO:OTIM?
'queries the operation hours.
Response: 100023
'the instrument was operated for 100023 hours up to now.
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic:INFO:PCOunter?
The command queries the number of power-on events.
Example:
DIAG:INFO:POC?
'queries the number of power on events.
Response: 123
'the instrument was switched on for 123 times up to now.
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic:POINt:CATalog?
The command queries the test points available in the instrument. A detailed description of the test
points can be found in chapter 3 , section "Trouble Shooting" of the Service Manual (on CD-ROM,
supplied with the instrument).
Example:
DIAG:POIN:CAT?
'queries the test points available in the instrument.
Response:
'DIAG_SMB_BAB_P7V,DIAG_SMB_BAB_P5V,DIAG_SMB_BAB_P3V3,…'
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic[:MEASure]:POINt?
The command triggers voltage measurement at the specified test point and returns the measured
voltage. A description of the test points can be found in chapter 3 , section "Trouble Shooting" of the
Service Manual (on CD-ROM, supplied with the instrument).
A list of the available test points can be queried using the DIAG:POIN:CAT? command .
Example:
DIAG:POIN? 'DIAG_SMB_BAB_P7V'
'triggers measurement at the test point DIAG_SMB_BAB_P7V.
Response: 0.5
'the voltage at the test point is 0.5 volt.
*RST value Resolution Options SCPI
- - Device-specific