Command Description R&S SMB
1407.0806.32 6.10 E-2
DISPlay Subsystem
The DISPlay system contains the commands to set the power-save mode of the instrument.
Command Parameters Default
unit
Remark
DISPlay:PSAVe:HOLDoff 1…60
DISPlay:PSAVe:STATe ON | OFF
DISPlay:PSAVe:HOLDoff 1 … 60
This command sets the wait time for the screen-save mode of the display. The available value range
is 1 to 60 minutes, the resolution 1 minute. The entry is dimensionless.
Example:
DISP:PSAV:HOLD 10
''sets the wait for the screen saver mode to 10 minutes.
*RST value Resolution Options SCPI
1 - Device-specific
DISPlay:PSAVe[:STATe] ON | OFF
This command activates/deactivates the screen-save mode of the display. With the screen-save
mode activated the display including backlight is completely switched off after the elapse of the wait
time (see command DISPlay:PSAVe:HOLDoff) when no entries via front panel, external mouse or
external keyboard are made.
This mode is recommended for preserving the display especially if the instrument is exclusively
operated via remote control.
Example:
DISP:PSAV ON
''activates screen saver mode.
*RST value Resolution Options SCPI
OFF - Device-specific