Command Description R&S SMB
1407.0806.32 6.20 E-2
OUTPut Subsystem
The OUTPut system contains the commands which set the properties of the RF output connector. The
properties of the LF output connector are set in the SOURce:LFOutput system, see section
"SOURce:LFOutput Subsystem".
Command Parameters Default
unit
Remark
OUTPut:AFIXed:RANGe:LOWer?
dBm Query only
OUTPut:AFIXed:RANGe:UPPer?
dBm Query only
OUTPut:AMODe AUTO | FIXed
OUTPut:BLANk:LIST:STATe ON | OFF
OUTPut:IMPedance
Ohm Query only
OUTPut:PROTection:CLEar
No query
OUTPut:PROTection:TRIPped
Query only
OUTPut[:STATe] ON | OFF
OUTPut[:STATe]:PON OFF | UNCHanged
OUTPut:AMODe AUTO | NORMal
The command switches the mode of the attenuator at the RF output (A
ttenuator MODe).
Parameters: AUTO
The attenuator is switched automatically. The level settings are made in the
full range..
FIXed
The level settings are made without switching the attenuator. When this
operating mode is switched on, the attenuator is fixed in its current position
and the resulting variation range is defined.
Example:
POW:ALC ON
'activates automatic level control for RF output.
OUTP:AMOD FIX
'sets the fixed mode with uninterrupted level for the RF output'.
*RST value Resolution Options SCPI
AUTO - Device-specific
OUTPut:AFIXed:RANGe:LOWer?
The command queries the minimum level which can be set without the attenuator being adjusted
(A
ttenuator FIXed).
The command is a query and therefore has no *RST value.
Example:
OUTP:AFIX:RANG:LOW?
'queries the minimum level for the FIXed setting'.
Response:
'-50' 'the minimum level is -50 dBm.
*RST value Resolution Options SCPI
- - Device-specific