R&S SMB Command Description
1407.0806.32 6.21 E-2
OUTPut:AFIXed:RANGe:UPPer?
The command queries the maximum level which can be set without the attenuator being adjusted
(A
ttenuator FIXed).
The command is a query and therefore has no *RST value.
Example:
OUTP:AFIX:RANG:UPP?
'queries the maximum level for the FIXed setting for the RF output'.
Response:-27
'the maximum level is -27 dBm.
*RST value Resolution Options SCPI
- - Device-specific
OUTPut:IMPedance
The command queries the impedance of the RF output. This permits converting the output level
between units V and W. The impedances cannot be changed.
The command is a query and therefore has no *RST value.
Example:
OUTP:IMP?
'queries the impedance of the RF output.
Response: 50'the impedance is 50 ohms
*RST value Resolution Options SCPI
- - Compliant
OUTPut:BLANk:LIST:STATe ON |OFF
The command activates/deactivates RF output blanking.
The command is a query and therefore has no *RST value.
Example:
":OUTP:BLAN:LIST:STAT ON
'activates the RF Output blanking.
*RST value Resolution SCPI
ON - Device-specific
OUTPut<[1]|2>:PROTection:CLEar
The command resets the protective circuit after it has been tripped. The state of the output is again
determined by OUTPut:STATe. .
The output is protected by a protective circuit which deactivates the output in the case of an
externally applied overvoltage This does not change the value of OUTPut:STATe.
Example:
OUTP:PROT:CLE" 'resets the protective circuit for the RF output.
*RST value Resolution Options SCPI
- - Compliant