EasyManua.ls Logo

Rohde & Schwarz SMB 100a - Page 214

Rohde & Schwarz SMB 100a
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Description R&S SMB
1407.0806.32 6.22 E-2
OUTPut:PROTection:TRIPped
The command queries the state of the protective circuit.
The command triggers an event and therefore has no reset value.
Example:
"OUTP:PROT:TRIP?" 'queries the state of the protective circuit for RF output A.
Response:"0" 'the protective circuit has not tripped.
Response:"1" 'the protective circuit has tripped.
*RST value Resolution Options SCPI
- - Compliant
OUTPut[:STATe] ON | OFF
This command activates and deactivates the RF output.
Example:
OUTP OFF
'deactivates the RF output.
*RST value Resolution Options SCPI
- - Compliant
OUTPut[:STATe]:PON OFF | UNCHanged
This command selects the state which the RF output assumes when the instrument is switched on.
The command is an event and therefore has no *RST value and no query form.
Parameters: OFF
The output is deactivated when the instrument is switched on.
UNCHanged
When the instrument is switched on, the output remains in the same state as it
was when the instrument was switched off.
Example:
OUTP:PON OFF
'The RF output is deactivated when the instrument is switched on.
*RST value Resolution Options SCPI
- - Device-specific

Table of Contents