R&S SMB Command Description
1407.0806.32 6.89 E-2
SYSTem:COMMunicate:GPIB:LTERminator EOI | STANdard
The command sets the terminator recognition for remote control via the IEC/IEEE bus.
Parameters: EOI
The terminator must be sent together with the line message EOI (End of Line).
This setting is recommended for binary block transmissions where a character
could coincidentally have the value LF (Line Feed) but is not intended as the
terminator. This setting must be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of whether it is
sent with or without EOI.
Example:
SYST:COMM:GPIB:LTER EOI
'only a character which is sent simultaneously with the line message EOI is
accepted as the terminator.
*RST value Resolution Options SCPI
STANdard - Compliant
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 1 ... 30
The command sets the IEC/IEEE-bus address .
Example:
SYST:COMM:GPIB:ADDR 14
'sets IEC/IEEE-bus address 14.
*RST value Resolution Options SCPI
28 - Compliant
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname ‘string’
The command enters the individual computer name of the R&S SMB. The entry is only activated
after sending command SYST:COMM:NETW:COMM:SET.
Note:
The hostname can only be changed after deactivating protection level 1 (command
SYSTem:PROTect<n>:STATe OFF, password)
Example:
SYST:COMM:NETW:HOST ‘SMB’
'enters the individual computer name of the R&S SMB
SYST:COMM:NETW:COMM:SET ON
'activates the change of the hostname.
*RST value Resolution Options SCPI
- Device-specific