Command Description R&S SMB
1407.0806.32 6.96 E-2
SYSTem:KLOCk ON | OFF
The command (K
eyboard LOCk) disables the front panel keyboard of the R&S Signal Generator
including the
LOCAL key, or enables it again (OFF).
Example:
SYST:KLOC ON
'activates the keyboard lock. The keyboard cannot be operated again until it
has been enabled with SYST:KLOC OFF.
*RST value Resolution Options SCPI
OFF - Compliant
SYSTem:PRESet
The command triggers an instrument reset. It has the same effect as the
PRESET key on the front
panel and the *RST command.
The command triggers an event and therefore has no *RST value and no query form.
Example:
SYST:PRES
'all instrument settings (also those that are not currently active) are reset to
their default values.
*RST value Resolution Options SCPI
- - Compliant
SYSTem:PROTect<n>[:STATe] ON | OFF, password
The command activates and deactivates the specified protection level. There are several protection
levels which disable specific service functions (authorized personnel of R&S Service Departments
only). These levels are identified by the suffix under PROTect.
The respective functions are disabled when the protection level is activated. No password is
required for activation. A password must be entered to deactivate the protection level. The
password for the first level is 123456. This protection level can be used to lock-out internal
adjustments.
The command triggers an event and therefore has no *RST value and no query form.
Example:
SYST:PROT1 ON
'activates protection level 1. Internal adjustments are only possible after
deactivating the lock-out.
SYST:PROT1 OFF, 123456
'deactivates protection level 1. Internal adjustments are enabled again.
*RST value Resolution Options SCPI
- - Compliant