R&S SMB Command Description
1407.0806.32 6.97 E-2
SYSTem:SERRor?
This command returns a list of all errors existing at the time when the query is started. This list
corresponds to the display on the info page under manual control.
The command is a query command and therefore has no *RST value.
Example:
"SYST:SERR?"
'queries all errors existing in the error queue.
Response:
"-221, 'Settings conflict', 153, 'Input voltage out of
range'"
'the two returned errors have occurred since the error queue was last queried.
*RST value Resolution Options SCPI
- - Device-specific
SYSTem:TIME 0...23,0...59,0...59
The command sets the time for the instrument-internal clock.
Example:
SYST:TIME 12,0,0
'sets the time to precisely 12 pm.
*RST value Resolution Options SCPI
- - Compliant
SYSTem:VERSion?
The command queries the SCPI version with which the instrument complies.
The command is a query command and therefore has no *RST value.
Example:
SYST:VERS?
'queries the SCPI version.
Response: 1996
'the instrument complies with the version from 1996.
*RST value Resolution Options SCPI
- - Compliant