Command Description R&S SMB
1407.0806.32 6.98 E-2
TEST Subsystem
The TEST system contains the commands for performing the routines as well as for direct manipulation
of the hardware assemblies (:TEST:DIRect). The self tests return a "0" if the test is performed
successfully, otherwise a value other than "0" is returned. None of the commands of this system have
an *RST value.
Caution!
The respective hardware assembly responds directly to the :TEST:DIRect
command; any safety mechanisms are bypassed. The command is used for
servicing purposes and should not be applied by the user. Improper use could
destroy the assembly.
Command Parameters Default
unit
Remark
TEST:DIRect 'assembly, subaddress, hex data string
TEST:DIRect ‘assembly', subaddress, hex data string
The respective hardware assembly responds directly to the command; any safety mechanisms are
bypassed. This function is only available via remote control.
Example
:
TEST:DIR 'SSYN',0,#H12345678
TEST:DIR? 'SSYN',0
Response: '#H12345678
*RST value Resolution Options SCPI
- - Device-specific