R&S SMB Command Description
1407.0806.32 6.101 E-2
SINGle
One complete sweep cycle is triggered by the IEC/IEEE-bus commands
...:SWEep:FREQ:EXEC or *TRG or TRIG:FSW:IMM. The mode has to be set
to AUTO (SOURce:SWEep:FREQ:MODE AUTO).
EXTernal
The sweep is triggered externally via the INST TRIG connector.
EAUTo
The sweep is triggered externally via the INST TRIG connector. As soon as
one sweep is finished, the next sweep is started. A second trigger event stops
the sweep at the current frequency, a third trigger event starts the trigger at
the start frequency, and so on.
Example:
TRIG:FSW:SOUR EXT
'selects triggering with an external trigger. The trigger is input via the INST
TRIG connector.
*RST value Resolution Options SCPI
SINGle - Device-specific
TRIGger[:PSWeep][:IMMediate]
The command immediately starts an RF level sweep. The command is only effective for sweep
mode Single (SOURce:SWEep:POWer:MODE AUTO in combination with TRIG:PSW:SOUR SING).
The command corresponds to the manual-control command Execute Trigger.
This command triggers an event and therefore has no *RST value.
Example:
SWE:POW:MODE AUTO
'selects the triggered sweep mode, i.e. a trigger is required to start the sweep.
TRIG:PSW:SOUR AUTO
'sets the Single trigger mode, i.e. a trigger starts a single sweep.
TRIG:PSW
'starts a single RF level sweep.
*RST value Resolution Options SCPI