Command Description R&S SMB
1407.0806.32 6.102 E-2
TRIGger:PSWeep:SOURce AUTO|IMMediate | SINGle | EXTernal|BUS | EAUTo
The command sets the trigger source for the RF level sweep.
The names of the parameters correspond directly to the various settings under manual control. SCPI
uses other names for the parameters; these names are also accepted by the instrument. The SCPI
names should be used if compatibility is an important consideration. An overview of the various
names is given in the following table:
R&S name SCPI name Command under manual control
AUTO IMMediate Auto mode
SINGle BUS Single mode.
EXTernal EXTernal Ext Single and Ext Step mode. The
command :SWEep:POW:MODE is used
to select between the two sweep modes.
Parameters:
AUTO | IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled continuously. As
soon as one sweep is finished, the next sweep is started.
SINGle
One complete sweep cycle is triggered by the IEC/IEEE-bus commands
...:SWE:POW:EXEC or *TRG or TRIG:PSW:IMM. The mode has to be set to
AUTO (SOUR:SWE:POW:MODE AUTO).
EXTernal
The sweep is triggered externally via the INST TRIG connector.
EAUTo
The sweep is triggered externally via the INST TRIG connector. As soon as
one sweep is finished, the next sweep is started. A second trigger event stops
the sweep at the current level, a third trigger event starts the trigger at the
start level, and so on.
Example:
TRIG:PSW:SOUR EXT
'selects triggering with an external trigger. The trigger is input via the INST
TRIG connector.
*RST value Resolution Options SCPI
SINGle - Device-specific
TRIGger[:SWEep][:IMMediate]
The command starts the sweeps which is activated. The sweep to be executed depends on the
respective MODE setting (:SOUR:SWEep:POW|FREQ:MODE and
:SOUR:LFO:SWEep[:FREQ]:MODE). The command corresponds to the manual-control command
Execute Trigger.
This command triggers an event and therefore has no *RST value.
Example:
TRIG 'starts all active sweeps.
*RST value Resolution Options SCPI
- - Device-specific