Example of Setup R&S SMB
1407.0812.62 2.10 E-2
Switch on RF frequency sweep by selecting State
On.
Now the Execute Single Sweep button to trigger
the single sweep appears.
The RF is not yet active, which means that no RF
signal is output.
Step 3: Activate RF signal
Active RF output by pressing the
RF ON/OFF key.
To indicate the active state, the RF block is
displayed in blue. An RF signal with the default
frequency and level settings is output, i.e. 1 GHz
and - 30 dBm.
The single sweep still must be triggered in the
sweep menu
Step 4: Trigger single RF frequency sweep
Select the Execute Single Sweep button in the
RF Freq Sweep menu by turning the rotary knob.
Press the rotary knob.
A linear single sweep from 100 MHz to 500 MHz in 1 MHz steps is output with a dwell time of 10 ms per
step.