R&S SMC100A Command Description
1411.4060.32 6.29 E-1
SENSe[:POWer]:STATus[:DEVice]?
The command queries if a sensor is connected to the signal generator.
The command is a query command and therefore has no *RST value.
Example:
SENS:STAT?
'queries if a sensor is connected.
R
esponse: 1
'a sensor is connected.
*RST value Resolution SCPI
- - Device specific
SENSe[:POWer]:SVERsion?
The command queries the software version of the connected R&S NRP power sensor.
The command is a query command and therefore has no *RST value.
Example:
SENS:POW:SVER?
'queries the software version of the R&S NRP power sensor.
*RST value Resolution SCPI
- - -
SENSe[:POWer]:TYPE?
The command queries the type of sensor . The type is automatically detected.
The command is a query command and therefore has no *RST value.
Example:
"SENS:TYPE?"
'queries the type of sensor.
Response: NRP-Z21
'the R&S NRP-Z21 sensor is used.
*RST value Resolution SCPI
- - Device specific
SENSe[:POWer]:ZERO
The command activates the autozero function. Zeroing is required in regular interval (at least once a
day)and if the temperature has varied more than about 5 °C, if the sensor has been replaced or if
measurements of signals with very low power are to be performed. The sensor must be
disconnected from all power sources before starting the autozero function.
The command is an event and therefore has no *RST value.
Example:
SENS:ZERO
'activates autozero function.
*RST value Resolution SCPI
- - Device specific