1. Daily Care, Inspection and Cleaning
7-1
This chapter explains how to maintain the system. So that the system operates properly, it must be maintained
correctly. Before performing maintenance, read this chapter carefully and follow the specified procedures.
To maintain this system in a usable condition, it must be maintained correctly and periodically. The table below
summarizes the recommended periodic maintenance items and their recommended frequencies. Those items
whose recommended frequency is "On condition" should be checked visually or via function check whenever
necessary. (Depending on the condition in which the system is used, some periodic maintenance actions may be
required more frequently than recommended.)
Clean the reagent bottles.
Cleaning the reagent bottle rack.
Clean the paraffin containers.
Clean the LCD protection sheet.
Clean the exterior panels and top cover of the instrument.
Clean the paraffin oven and reagent bottle spill tray.
Replace the activated carbon filter.
Replace the retort lid gasket.
In addition to the periodic maintenance performed by the user, some of the key system and parts must be inspected
and replaced periodically. The recommended periodic inspection/replacement tasks are summarized in the table
below, together with the recommended frequencies.
The following replacement tasks must be performed by the service personnel. Contact our Customer Service
Department or our local representative.
Replace the air pump diaphragm.
On condition at periodic inspection
Replace O-rings in the rotary valve.
On condition at periodic inspection
3) Recommendation of Warm Water Flush to Users of Neutral Buffered Formalin
Warm water flush is performed to wash away with warm water the precipitate build-up that forms in the tubing when
fixing agent with buffered agent (neutral buffered formalin) is used. Precipitate build-ups between the respective
stations and the retort are removed. (Warm water flush: p. 3-2)