1. Installation Method
1-1
1) Installation of the Instrument
This section provides information on determining a location for, and installing the instrument.
Installation should be performed by qualified service staff working exclusively with the instrument. The instrument
must be installed correctly to ensure proper operation and service.
Read this Operating Manual carefully before operating the instrument. Strictly observe all instructions provided in
this document.
Caution!
The instrument is a precision instrument and must be handled with care. If the
instrument is handled in a rough manner or dropped, its internal components may
malfunction or get damaged. Always exercise due care when handling the
instrument.
[1] Flow of installation
Details: p. 1-2
Checking the installation conditions
Checking the work area
Transporting the instrument
Details: p. 1-4
Unpacking and installation
Installation check
Details: p. 1-13
Turning on the power
[2] Required tools
Name
Size
Q'ty
Nut screwdriver
10 mm across flats
1
Phillips screwdriver
ISO No. 2
1
Level
-
1
Convex ruler (steel tape measure)
Capable of measuring objects of
2.5 m or larger
1
[3] Installation environment
As is the case with all electronic devices, the instrument should not be installed in places subject to excessive
temperature or humidity.
Temperature and humidity should be held at constant levels to achieve the highest degree of operating stability.
The ambient operating temperature of this instrument is in a range of +10 to +40C. Its ambient operating humidity
is in a range of 30 to 85% in relative humidity (non-condensing). Also, locate the instrument, with at least 10 cm
away from walls, in a well-ventilated place not exposed to corrosive vapor or extremely high/low temperatures or
where the instrument will not come in direct contact with air flow. Install the instrument away from direct sunlight
and do not install it proximity to open windows, sinks, ovens, open flames, hot plates, radiators, and dry ice baths.
The instrument should be installed away from any equipment carrying high voltage or current, such as a large
refrigerator or oven. Since the instrument is very heavy, the floor should have sufficient bearing strength. Install the
instrument on such floor that has been laid flat and the instrument should be secured with caster lockdown feet.
Also install anti-tipover brackets, etc., in case of earthquake. The instrument should be installed near a power
supply meeting the power facility requirements (voltage, current) specified in this Operating Manual. Be sure to
ground the power outlet and wire the power properly using a dedicated cable.
warning!
If the instrument is used in a non-ventilated place, serious health risks may occur to the human body.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals