1. Installation Method
1-9
Space for spill tray
Spill tray
Space for spill tray
Spill tray
8) Checking the Installation (Securing the Instrument)
1. Confirm that the location where the instrument will be installed is flat and sufficiently strong.
2. Move the instrument to the installation location and lower the stoppers of the front casters of the instrument to
secure the instrument.
If enough space is left for people to pass through at the rear of the instrument, the users may touch the power
cord or hook it with their foot and accidentally cut off the power; accordingly, select a location where no
unnecessary large space will be left at the back of the instrument.
9) Installation Check (Installing the Accessories)
[1] Installing the spill tray (for paraffin oven)
Put the spill tray in the space below the paraffin oven and push it all the way in, as shown in the figure below.
[2] Installing the spill tray (for Reagent bottle rack)
Put the spill tray in the space below the Reagent bottle and push it all the way in, as shown in the figure below.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals