Touching the Parts Usage icon on the History tab screen displays the Parts Usage screen.
On the Parts Usage screen, you can check or output the usage statuses of parts. If the Current Usage value
exceeds the Limit value, it means the part must be replaced.
[1] Parts usage display field
The name of a part is displayed.
The unit in which the part is managed is
displayed.
• Pump diaphragm
Time (Operating hours of the pump)
• Rotary valve maintenance
Runs (Counted up every time tissue processing is
performed)
• Retort seal
Days (Counted up every time the date changes)
• Activated carbon filter
Days (Counted up every time the date changes)
The status of usage is shown based on the Unit
field.
The preset limit is shown.
[2] Clear button
Touch and select the Current Usage field of the part
whose usage data you want to clear, and then touch
Clear to clear the usage data.
Usage data can be cleared only for the retort seal and
activated carbon filters.
If the replacement timing is passed, appears in the
bottom right-hand corner of the standby screen.
Touch , and clear the usage data on the main screen
after checking the details. Should you encounter any
situation you cannot handle, Contact the Sakura Finetek
Technical Support representative or local representative.
On the History tab screen, touch the Software Version icon to open the Software Version screen.
The software version and serial number of the instrument will be displayed.