On the Utility tab screen, touch the Edit Programs icon to
open the Select Program screen.
Select the program you want to edit, and touch Edit.
The Edit Program screen appears. On the Edit Program
screen, "Time," "Temp," "P/V," "Mix" and "Predicted
Endtime" can be set. You can also set these items
individually or for each station.
If the program currently used is selected, a
confirmation message appears.
The program cannot be saved with all stations
set to "0" processing time.
If the processing time of a station for which
P/V or temperature is set is "10 hours or
more," a warning window appears. Check the
following precaution.
Very volatile solutions evaporate easily when
the pressure is reduced, so avoid setting the
P/V to "ON" for these solutions. Errors and
tissue processing failures may occur.
[1] Program details display field
Program details are displayed.
By touching the "Sta." or "Solution Name" field, you
can copy the settings of "Time," "Temp," "P/V," "Mix"
and "Predicted Endtime" for the selected station, to
all other stations.
You can touch the "Time," "Temp (C)," "P/V" and
"Mix" fields to set these items individually.
[2] Predicted Endtime field
You can set a predicted endtime.
[3] Editing the program processing time
In the program details display field, touch Time [1] for
the station you want to edit, and a value entry window
will appear.
Enter a value in a range of 0:00 to 99:59.
Touching Enter saves the settings, after which the
previous screen will be displayed again.