1. Daily Care, Inspection and Cleaning
7-4
7) Cleaning of Exterior Panels
Protect the exterior panels from attachment of dust and dirt. In general, dust attached to the exterior panels can be
removed only by wiping with a clean cloth wrung tightly after moistened with water.
To clean the front surface of the monitor screen, wipe with 10% diluted ethanol solution and then let the surface dry
naturally.
Wipe the reagent bottle rack door using glass cleaner and a soft cloth. If paraffin is attached, wipe it off with a cloth
moistened with Tissue Clear (or xylene).
Caution!
Unless otherwise specified in this manual, never use solvent of any kind for cleaning, etc.
[1] Cleaning of the screen protection sheet
If the protection sheet is dirty, remove the sheet and rinse it with water. Replace the protection sheet if badly
scratched, etc.
8) Cleaning of Paraffin Containers
Clean the paraffin containers as necessary.
Warning!
The paraffin oven and paraffin may be hot, so be careful not to burn yourself.
1. Open the door for the paraffin oven and remove the paraffin container to be cleaned.
2. Empty the paraffin container.
3. Use a paper towel to wipe the interior and exterior of the paraffin container. Use a paraffin scraper to clean, if
necessary.
4. Return the paraffin container to its original position and close the door for the paraffin oven.
9) Cleaning of the Retort
Clean the retort as necessary.
Caution!
If buffered formalin is used, white precipitates may deposit in the retort. In this case, clean the interior walls
of the retort using a cloth moistened with cold or warm water. Thereafter, wipe dry the walls with alcohol and
let them dry.
1. Open the retort lid. If the retort is hot, wait until it becomes cool.
2. Wear gloves and wipe the interior surface of the retort, lid gasket and gasket-contacting surface using a cloth
moistened with Tissue Clear (or a small amount of xylene).
3. Check the retort lid gasket for scratches. The gasket with scratch marks must be replaced.
4. Remove the strainer installed at the bottom of the retort.
To remove the strainer, turn the screw counterclockwise.
Clean the strainer using a brush or toothbrush.
Caution!
Prevent foreign matters from getting in the strainer when
removing from the retort.
5. Attach the strainer back to its original place and tighten the screw.
Figure of the retort
viewed from the top
Strainer
Front of instrument
Back of instrument
Screw

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals