1. Operating Procedure
2-13
6-
2-
6) Ending Tissue Processing
When all steps in the tissue processing program end, the
Tissue Processing completed window appears.
Since you as the operator are currently logged off, log on
first and then remove the specimens and drain the paraffin
from the retort.
When finished, proceed to retort cleaning.
Note
If the backlight is off, it turns on automatically.
[1] Removing the specimens and draining the
paraffin
Once the password is accepted in the Tissue Processing
completed window, this window appears. Remove the
specimens and drain the paraffin from the retort according
to the message. You can perform the two operations in any
order.
Removing the specimens
1. Touch Take out basket.
2. Follow the messages to open the retort lock cover, remove the specimens from the retort, and touch Yes.
Close the retort lid and close the retort lock cover, and touch START to drain retort.
3. When draining is complete, the retort cleaning start screen appears.
Draining retort first
1. Touch Drain retort.
2. When draining is complete, follow the message to open the retort lock cover.
3. Remove the specimens from the retort and click Yes to open the Clean Retort screen. The retort cleaning
start screen appears.
Draining

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals