1. Operating Procedure
2-10
Delayed Processing Screen
Tissue Processing Screen
[3] Monitoring tissue processing
Touching START in the window to confirm start of tissue processing starts tissue processing.
If "Immediate Start" is selected as the tissue processing starting method, the Tissue Processing screen appears.
If "Delayed Start" is selected, the instrument will stand by until it is time to start processing.
Caution!
Check the solution levels before starting the operation.
During tissue processing, make sure the paraffin oven door is securely closed. If the door is open, paraffin
is not filled/drained to ensure safety.
The solution levels in the illustration are not accurate. They may differ from the actual solution levels.
The following operations are permitted during delayed processing and tissue processing.
Delayed Processing
Pauses the instrument, and immediately start tissue processing.
Pauses the instrument, and abort tissue processing.
Pauses the instrument, and change the current program.
Pauses the instrument, and resume tissue processing (delayed start).
Pauses the instrument, and check if paraffin has melted.
Tissue Processing
Pauses the instrument, and resume tissue processing.
Pauses the instrument, and abort tissue processing.
Pauses the instrument, and change the current program.
Pauses the instrument, and check if paraffin has melted.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals