[4] Setting a usage limit for solution
Touch the area where the solution usage limit is
displayed [1], and an input limit window will be displayed.
Enter a value according to the selected management
method.
Touching Save saves the settings, after which the Utility
screen will be displayed.
Manage the solution exchange frequency in
days (1 to 99 days).
Factory default setting: 5
Manage the solution exchange frequency in
runs (1 to 99 runs).
Factory default setting: 5
Manage the solution exchange frequency in
cassettes (1 to 9999 cassettes).
Factory default setting: 1500
Using solutions more often may affect the
specimens quality. Using cleaning solutions
more often may cause cleaning problems or
instrument failures.
[5] Setting when to perform warm water flush
If reagents such as buffered formalin are used,
precipitates are generated and may clog the tubing.
Accordingly, warm water flush is performed regularly and
when to perform warm water flush is managed.
Touch the setting area [1] in the Flush field displays a
value entry window, so enter a warm water flush
frequency.
Touching Save saves the settings, after which the Utility
screen will be displayed.
Not Managed (Factory default setting)
If the number of runs reaches the specified
number of times, a message urging warm water
flush appears.
Increasing the number of tissue processing
before warm water flush is performed may
cause cleaning problems or instrument
failures.
The message appears before the retort
cleaning completion window and the tissue
processing preparation step.
Refer to p. 3-2 for information on how to
perform warm water flush.