EasyManuals Logo

Sakura Histo-Tek VP1 User Manual

Sakura Histo-Tek VP1
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
1. Installation Method
1-10
[3] Installing the Divider for fume control unit and
activated carbon filters
Warning!
The fume control unit reduces the amount of organic solvent released from the instrument. To make sure
fume is adsorbed and treated properly, replace the activated carbon filters in the fume control unit
periodically.
So that the gases generated during tissue processing are filtered properly, keep the activated carbon filters
always installed.
1. Open the fume control unit door. Push in the bottom
of the door to unlock and open the door.
2. Put the Divider for fume control unit into the
instrument.
Set the unit so that the side with the larger hole
comes to the front. Make sure the front of the unit is
inside the instrument. If the front side is only resting
at the entrance, fume cannot be treated properly.
3. Install the activated carbon filters to the instrument.
Two activated carbon filters constitute one set.
Fume cannot be treated properly if only one filter is
set.
4. Close the fume control unit door. Push in the door
until you hear a click sound, to lock the door.
The side with the larger
hole comes to the front.
Fume control unit door
The front of the base is
inside the instrument:
Correct setting
Fume control unit door
Activated carbon filters
The front of the base is
resting at the
entrance: Wrong
setting
Divider for fume
control unit

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals