§ Press the number key corresponding to the content to
be added to the Result Printout format�
§ The following content is either On or Off:
Analyzer info Operator
Calibration info Program name
Program info Weight info
Signature line
§ The following content requires a selection or input:
Header
Sample number
Interval print
A complete printout format is shown to the left:
50
Header line 1
Header line 2
Header line 3
LMA100P Moisture Analyzer
Module 1: M30604026
Operator: John Smith
Sample ID: 1
Calibration internal: 06/12/2012 09:00 AM
Last temp cal: 05/13/2003 08:20 AM
Program 1: Titon X
UNITS: Moisture FACTOR: 1.00, 0.00
MODE: Standard START DELAY: 9 sec M2
TEMP1: 105C, Std TIME1: Off
TEMP2: Off TIME2: Off
SLOPE: Window: 1.0 %IW: 0.050 Actual
STANDBY: 60C EQUILIBRATE: Off*
IDEAL WT: 5.0 grams LOCK: Off
TARGETS: MIN: 45.00 MAX: 55.00
Time Temp Weight Data Units
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
0:00 61C 5.167 0.00 %M
RESULT on 06/15/2012 at 10:24 AM
Elapsed time: 07:13
52.466 %M
Target: PASS
Initial Weight = 5.1671 grams
Final Weight = 2.4561 grams
Weight Loss = 2.7110 grams
Signature:
* = only for model LMA100P

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Mark 3 and is the answer not in the manual?

Sartorius Mark 3 Specifications

General IconGeneral
BrandSartorius
ModelMark 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals