76
§  The automatic temperature adjustment report is printed 
after the adjustment process�
  To verify an adjustment temperature:
§  From the Service menu, press the 3 key to display the 
Adjust Temperature screen�
§  Press the 2 for Verify temperature� The screen will 
change to the Verify temperature screen�
§  The default verifies temperature is 130°C� Press the 
“Temp” soft key to begin� The screen will change to the 
Temp verify in process screen and show the selected 
temperature� After 20 minutes, the screen will change 
to the Temp verify done screen and will display the 
expected selected temperature and the actual tempera-
ture that was measured by the  
temperature bowl (“fixture”)�
§  Press the Print soft key to print a temperature  
verification report (LMA100P only)�
§  If the agreement of the selected temperature and the 
actual temperature are not within 2 degrees for any of 
the selected temperatures, it is recommended that you 
perform the temperature calibration and adjustment 
procedure�
Automatic Temperature Adjustment Report 
- - - - - - - - - - - - - - - - - - - - - - - - - -
Fixture number 1234 
Temperature  80C:    82C 
Temperature   150:    150C 
Mark 3 Moisture Analyzer 
Module SN: M3085010B 
Calibration passed 04/10/11  02:05 PM 
Signature: 
__________________________ 
              
Preadjust 80: 79
Adjusted 80:  80
Preadjust 150: 144
Adjusted 150:  150
Temp adjust done 
Print
Ad-