To change the temperature adjustment schedule:
§ From the Service menu, press the 3 key to display the
Adjust Temperature screen�
§ Press the 3 for Adjust schedule� The screen will change
to the Schedule temp adjust screen�
§ Select the desired period of time� When this period
expires, a warning will appear in the warning box
indicating that a new verification or temperature
adjustment should be performed�
To enter a reference fixture number:
§ From the Adjust Temperature screen, press the 4
for Reference fixture� The screen will change to
have you enter the reference fixture number that is
marked on your temperature bowl� The number will
then appear on both the temperature adjustment or
verification reports�
1) Off
2) 1 day
3) 7 days
4) 30 days
5) 90 days
6) 180 days
7) 365 days
Schedule calibration
09/07/2012 02:14 PM
Calibration schedule: Off