Parameters
INTECONT® Tersus Belt weigher, Instruction Manual
Schenck Process Group
BV-H2464GB, 1234
- 65 -
P02.04 Pulses per length
Default: 50.26 I/m
Min: 0.1 I/m Max: 1000000 I/m
Number of speed transducer pulses per belt length meter. The parameter is important for
the accuracy of the scale.
If noch tacho is active don't change characteristic value vs. This would change the timing of
calibration programs.
P02.05 Nominal Speed
Default: 1 m/s
Min: 0.001 m/s Max: 50 m/s
Serving as reference value for limit value monitoring, nominal speed determines nominal
belt load together with nominal feed rate.
Value should be accurate to some percent.
P02.06 Feeder Start
Default: OP
Alternative: OP
FB
DI
V
Specifies the device that will turn on the scales.
The digital input is flank-controlled, i.e. the ON signal must be removed and then reapplied
if the scales becomes inoperative because of e.g. an alarm.
P02.07 Counter 1 Unit
Default: ------- t
Alternative: ------- kg
-----.- kg
----.-- kg
---.--- kg
------- t
-----.- t
----.-- t
---.--- t
Display format for counter 1.