Overview VC 1100
Version 2 overview//VC1100E
1-8 C01 / C02 - C11 / C12 © Schenck VIBRO GmbH, D-64273 Darmstadt, November 97
The following table shows the settling times and the cycle times
(sum of settling and measuring time) for different functions and
setups of the Vibration Signal Conditioner. Different setups can be
used for channel A and B. The total cycle time is the sum of the
cycle times for channel A and B.
Activated Function Settling Time Cycle Time
High Pass Filter: 1 Hz 1.75 s 4.75 s
High Pass Filter: 3 Hz 1.00 s 4.00 s
High Pass Filter.: 10 Hz ISO 1.25 s 4.25 s
High Pass Filter: Special 1.75 s 4.25 s
Integrator 6.00 s 9.00 s
Linearization Circuit: 5.75 s 8.75 s
If more than one function has been selected, the function with the
longest settling time determines the total settling time.
2 Channel Monitor (Vibration and Bearing Condition)
Since VIBROCONTROL 1100 is equipped with a Bearing Condit-
ion Detector, the measurement of Bearing Condition is indepen-
dent of the vibration measurement.
The settling time is 2.75 s and the measuring time is 1.25 s.