VC 1100 List of Setup Parameters
VC1100E/paralist Version 2 →
© Schenck VIBRO GmbH, D-64273 Darmstadt, Juni 98 C01 / C02 - C11 / C12 5-9
Signal Detection Type Group J
General
Group J defines how the incoming vibration signal is processed
and displayed.
The signal processing block diagram shows the function of the
setup parameter and uses the following terms:
Vibration Signal : Output of the vibration transducer.
Measured Parameter : Vibration displacement, Vibration
velocity, Vibration acceleration.
Unit : Unit of the measured parameter.
VIBROCONTROL 1100 uses the
same unit for the limit values
Signal detection type : rms, zero-to-peak calculated (pc),
peak-to-peak calculated (ppc)
Fig. 5 - 3: Parameter Group J, Signal Detection Type
Measuring ranges
The measuring range depends on the selected pickup type, mea-
sured parameter, and signal detection type. The range is conti-
nuously adjustable within the minimum and maximum values
shown in the table.