Built-in Operating Panel and Display VC 1100
Version 2 built-in/VC1100E
4-18 C01 / C02 - C11 / C12 © Schenck VIBRO GmbH, D-64273 Darmstadt, November 97
Communication errors
Error
code
see page 20 Meaning (continued)
-55 w
Correct parameter specifier
received, but command cannot
be processed because the reque-
sted data are not available; e.g.
channel not active
-57 w
received data not accepted,
because:
a) data is not in the list of
choices.
b) number (INTEGER or
FLOATING POINT) is out of
specified range.
-58 w
Received unit of the pickup sensi-
tivity (I04) does not correspond to
the selected pickup (e.g. mV/g
and vibration velocity pickup).
-59 w
The sensitivity (I05) is too large
or too small.
-60 w
The selected measured parame-
ter (J03/J04) cannot be proces-
sedwith the selected pickup (I03);
e.g. vibration acceleration with
vibration velocity pickup.
-61 w
The unit (J05/J06) does not
correspond to the measured pa-
rameter (J03/J04); e.g. vibration
displacement cannot be measu-
red in g.
-62 w
The selected measuring range
(J09/J10; J13/J14) is too small
or too large.