Built-in Operating Panel and Display VC 1100
Version 2 → built-in/VC1100E
4-12 C01 / C02 - C11 / C12 © Schenck VIBRO GmbH, D-64273 Darmstadt, November 97
Test analog output 1.
Select voltages of 0, 2, 5, or 10 Volt by pressing
Press to exit. Press
to go to next test.
Test analog output 1.
Select currents of 0, 4, 12, or 20 mA by pressing
Analog output 2 is tested in the same manner.
Analog Output 2 Voltage
S07 0 V
2 V
5 V
10 V
Analog Output 2 Current
S08 0 mA
4 mA
12 mA
20 mA
Press to start the self-test. During the self-test, a count down
from 10 to 0 is displayed. OK will appear on the display if the test
is completed successfully.
The self-test does not suspend the monitoring mode.