List of Setup Parameters VC 1100
Version 2 parqalist/VC1100E
5-20 C01 / C02 - C11 / C12 © Schenck VIBRO GmbH, D-64273 Darmstadt, Juni 98
L 03 Analog Output 1 : Measured Parameter
Choises : vib A, BCU A,vib B, BCU B
Function : Select measured parameter sent to analog output 1.
vib A = Measured Parameter:
Vibration Channel A
BCU A = Measured Parameter:
Bearing Condition Channel A
vib B = Measured Parameter:
Vibration Channel B
BCU B = Measured Parameter:
Bearing Condition Channel B
L 04 Analog Output 2 : Measured Parameter
Choises : vib A, BCU A,vib B, BCU B
Function : Select measured parameter sent to analog output 2.
vib A = Measured Parameter:
Vibration Channel A
BCU A = Measured Parameter:
Bearing Condition Channel A
vib B = Measured Parameter:
Vibration Channel B
BCU B = Measured Parameter:
Bearing Condition Channel B
L 05 Analog Output 1 : Range Full Scale
Choises : 0.000 ... 999.0
Function : In connection with an indicator or a plotter, the ana-
log output can be scaled to an optional full scale
value within the present range limits.
L 06 Analog Output 2 : Range Full Scale
Choises : 0.000 ... 999.0
Function : In connection with an indicator or a plotter, the ana-
log output can be scaled to an optional full scale
value within the present range limits.