APPENDIX E – AL A RM S
79.
Incorrect connection of the
temperature sensor (sterilization
chamber) to the connector.
Contact the Technical Support Department
(see Appendix Z).
Temperature sensor short circuit
(sterilization chamber).
Incorrect connection of the
temperature sensor (steam
generator) to the connector.
Temperature sensor short circuit
(steam generator).
Incorrect connection of the
temperature sensor (heating
element) to the connector.
Temperature sensor short circuit
(heating element).
Incorrect connection of the
temperature sensor (chamber wall)
to the connector.
Temperature sensor short circuit
(chamber wall).
Intervention of the steam generator
safety thermostat.
Manually rearm the thermostat(s) located on the back of the device (see
the Chapter 3 - Product Overview)
Unscrew the black plastic protection cap, press the white button until
you hear a click and replace the cap.
Turn-off (RESET) and then turn-on the device.
If the problem persists contact the Technical Support Department
(see Appendix Z).
Intervention of the heating element
safety thermostat.
Heating or steam generator heating
element malfunction.
Presence of water or condensate in
the sterilization chamber.
Perform RESET following the instructions.
Carefully dry the inside of the sterilization chamber and start the program
again.
Do not put material impregnated with water, or liquids in general, in the
chamber.
Drain filter of the sterilization
chamber obstructed.
Clean or replace the drain filter.
(See Appendix C “Maintenance”).
Air leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again. If the gasket still leaks, replace the gasket.
Contact the Technical Support Department (see Appendix Z).
Problem in the plumbing circuit.
Water injection pump malfunction.
Contact the Technical Support Department (see Appendix Z).
Problem in the plumbing circuit.
Intervention of the steam generator
safety thermostat.
Manually rearm the thermostat(s) located on the back of the device (see
the Chapter 3 - Product Overview).
Unscrew the black plastic protection cap, press the white button until
you hear a click and replace the cap.
Turn-off (RESET) and then turn-on the device.
If the problem persists, contact the Technical Support Department (see
Appendix Z)
Heating element safety thermostat
intervened.
Contact the Technical Support Department (see Appendix Z).
Heating or steam generator heating
element malfunction.