EasyManuals Logo

Servomex SERVOPRO MultiExact User Manual

Servomex SERVOPRO MultiExact
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #126 background imageLoading...
Page #126 background image
Appendix MultiExact Gas Analyser
APPENDIX
A5 DISPLAY UNIT CONVERSION FACTORS
When you select display units as described in Section 5.10, you must ensure that you
also enter the correct units conversion factor, as shown in the table below:
To convert
from *
to
use the units
conversion factor
applicable
gas(es)
%
ppm
10000
any
ppm
%
0.0001
any
ppm
vpm
1
any
ppm
mg/m
3
1.2492
CO
"
"
1.9631
CO
2
"
"
1.4277
O
2
%
mg/m
3
12492
CO
"
"
19631
CO
2
"
"
14277
O
2
ppm
%LEL
0.0008
CO
%
%LEL
8
CO
%
mol/mol
0.01
any
ppm
mol/mol
#
#
*
Measurement default units.
Selected display units.
#
This conversion is not supported.
To return to the measurement default units, select the "off"
units selection
option and set the units conversion factor to "1": see Section 5.10.
120 05410001A / Revision 6