MultiExact Gas Analyser Appendix
APPENDIX
A9 PERFORMANCE DATA
A9.1 O
2
purity measurement
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
Display indication
Measured volume % O
2
Full scale range
0 to 100% O
2
Analogue output range
User selectable (minimum difference0.5%)
Resolution
0.01% O
2
Repeatability
< 0.01% O
2
Linearity
No measurable error
Intrinsic error (accuracy)
± 0.01% O
2
Zero drift per week
± 0.01% O
2
Span drift per week
± 0.02% O
2
Output fluctuation
± 0.01% O
2
Response time *
10 seconds
Flow effect
Flow driven
± 0.1% O
2
Pressure driven
± 0.1% O
2
Zero temperature coefficient
± 0.01% O
2
per 10 °C
Span temperature coefficient
± 0.01% O
2
or 0.1% of reading per 10 °C **
Pressure effects
< 0.003% O
2
#
*
T
90
at 200 ml min
-1
or 68.9 kPa gauge (7 psig, 0.48 bar gauge) sample gas supply
pressure.
For a 100 to 250 ml min
-1
change in sample gas flow rate.
For a 14 to 55 kPa (0.14 to 0.55 bar, 2 to 8 psi) change in sample gas inlet pressure.
#
For a 1% change in ambient pressure.
** Whichever is the greater.
05410001A / Revision 6 125