A9 PERFORMANCE DATA
A9.1 O
2
purity measurement
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
User selectable (minimum difference – 0.5%)
Intrinsic error (accuracy)
Zero temperature coefficient
Span temperature coefficient
± 0.01% O
2
or 0.1% of reading per 10 °C **
T
90
at 200 ml min
or 68.9 kPa gauge (7 psig, 0.48 bar gauge) sample gas supply
pressure.
For a 100 to 250 ml min
-1
change in sample gas flow rate.
For a 14 to 55 kPa (0.14 to 0.55 bar, 2 to 8 psi) change in sample gas inlet pressure.
For a 1% change in ambient pressure.
** Whichever is the greater.