MultiExact Gas Analyser 5 – Installation and set-up
5.2 Electrical Installation
5.2.1 Electrical Safety
WARNING
Ensure that the electrical installation of the analyser conforms with all
applicable local and national electrical safety requirements.
WARNING
Obey the safety instructions given below when you install the
analyser; if you do not, the analyser warranty may be invalidated, the
analyser may not operate correctly, or it may be damaged.
The following warnings must also be considered:
The electrical supply coupler or plug must be easily accessible for disconnection
from the electrical supply.
The electrical supply circuit must incorporate a suitable fuse or over-current
protection device, set to or rated at no more than 10 A.
Ensure that your electrical supply can provide the necessary maximum power
consumption: refer to Section 2.3.
All signal and electrical supply cables must be rated for temperatures of 70 °C or
higher.
Ensure that the cables that you connect to the analyser are routed so that they do
not present a trip hazard.
When you carry out insulation testing, disconnect all cables from the analyser.
Ensure the analyser is provided with a sound earth connection via the electrical
supply plug (Figure 2, item 8).
05410001A / Revision 6 27

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO MultiExact and is the answer not in the manual?

Servomex SERVOPRO MultiExact Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO MultiExact
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals