MultiExact Gas Analyser Appendix
A9.5 Performance data: 0-500ppm CO
2
measurement
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
The performance specification for 0-50ppm range is applicable when
transducer is calibrated with 50ppm CO
2
span gas.
Full scale range
Lower specified range
2
User selectable
(minimum difference – 10% of full range)
0-500ppm¹: ±0.5ppm (below 25ppm), <3% of
reading (25
-500ppm)
0-500ppm¹: ±0.5ppm (below 25ppm), ±3% of
reading (25
-500ppm)
Intrinsic error (accuracy)
0-500ppm¹: ±0.5ppm (below 25ppm), ±3% of
reading (25
-500ppm)
± 0.3 ppm ± 2% of reading (for the specified
range)
± 0.1 ppm or 1% of reading **
†
± 0.1 ppm or 1% of reading **
‡
± 3% of measurement or 0.25 ppm per 10 °C **
05410001A / Revision 6 129