MultiExact Gas Analyser 5 – Installation and set-up
5.5.7 Automatic validation/calibration timers
1. Return to the auto val screen
(Figure 30). Select 'Timer'.
2. Select 'On' or 'Off' for Timer.
Figure 43 – The auto val/cal
timer screen
3. Scroll to the auto val/cal start date
screen.
4.
Edit the displayed start date as
described in Section 4.10.
Figure 44 – The auto val/cal
start date screen
5. Scroll to the auto val/cal start time
screen.
6.
Edit the displayed start time as
described in Section 4.10.
Figure 45 – The auto val/cal
start time screen
7. Scroll to the auto val/cal repeat time
screen (see Figure 46).
8.
Edit the displayed repeat time as
described in Section 4.10.
Figure 46 – The auto val/cal
repeat time screen
If you set the repeat time to "000:00" (0 days, 0
hours), the automatic
validation/calibration will only be performed once, on the specified start date,
at the specified start time. You will therefore need to initiate any subsequent
automatic validation/calibration: see Section 6.2.1.
05410001A / Revision 6 49